Tap Density Tester FM-TDM-A100

Tap Density Tester FM-TDM-A100

Tap density tester FM-TDM-A100 is a single test station unit. It’s designed in accordance with USP I & USP II Tapped density Method. Measures tapped density, flowability of powder to optimize size of the packaging container. Tapped density is calculated in gm/ ml by dividing the sample weight by the final tapped volume.

  •  Test figure accuracy up to 4 decimal places
  •  Stores test data in shut down mode
  •  Highly advanced result interpretation
  •  Built-in printer to print test report
  •  Measures Hausner Ratio (TD/BD), Compressibility Index (TD-BD/TD × 100)
  •  No need to adjust an integer
  •  Automatically saves test result

Sr. No. Standards
1 ISO 3953: Metallic powder determination of tap density
2 USP Tapped Density Method II
3 USP Tapped Density Method I
4 ASTM B 527
5 DIN ISO EN 787/11
6 EP 2.9.15

Work station 1
Sample volume ≤ 250 ml
Sample weight ≤ 500 g
Number of Tapping 0 to 99999 (set 0 to output bulk density)
Frequency 300 rpm (continuous adjustable in 0 to 300 rpm)
Amplitude Maximum 15 mm ( 1 to 15 mm integer variable)
Stroke height Stroke height 3 mm
Strokes/ min 250 strokes/ min
Display LED display
Repeatability ≤ 1 %
Accuracy ≤ 1 %
Humidity ≤ 85 % RH
Motor torque 0.86 N. m
Power 40 W
Power supply AC 220 V± 10 %, 50 Hz

Used to test tapped density, simultaneously calculate bulk density, Hausner ratio across production, packaging, quality and research testing units in pharma, food, medicine, life science and other industries.

Accessories no. Name Details Qty.
1 Washer Polyurethane 4 each
2 Glass measuring cup 25 ml 1 each
3 Glass measuring cup 50 ml 1 each
4 Glass measuring cup 100 ml 2 each
5 Glass measuring cup 250 ml 1 each
6 Damping washer 1 mm thickness 2 each
7 Damping washer 2 mm thickness 2 each
8 Damping washer 5 mm thickness 2 each
9 Power cable 250 V, 10 A 1 each
10 Fuse 5 × 20 mm, 1 A 5 each
11 Silicone gasket 70 × 2 mm 1 each
12 Tum brush 1 each