Sample Type | Whole blood, serum, plasma, cerebrospinal fluid, dilute urine |
Sample Volume | 200 μL |
Measuring Time | ≤ 25s |
Analysis Method | Ion selective electrode (ISE) |
Measuring Parameters | K+, Na+, Cl-, iCa2+, nCa2+, TCa2+, TCO2, pH, AG |
Measuring Range |
K+ 0.5 to 20 mmol/L Na+ 15 to 200 mmol/L Cl- 15 to 200 mmol/L Ca2+ 0.1 to 6 mmol/L TCO2 2 to 70 mmol/L pH 4 to 9 |
Resolution |
K+ 0.01 mmol/L Na+ 0.01 mmol/L Cl- 0.01 mmol/L Ca2+ 0.01 mmol/L TCO2 0.1 mmol/L pH 0.01 |
Working Temperature | 10 to 40 ℃ |
Relative Humidity | ≤ 80% |
Atmospheric Pressure | 86 to 106 kPa |
Number Of Sample Positions | 30 |
Injection Mode | Manual |
Calibration | Automatic and Manual calibration |
Display | 7-inch HD touch screen |
Language | Customizable |
Storage | Up to 50,000 test results |
Printer | Built-in thermal printer |
Interface | RS232 port |
Power Supply | AC 110 to 240 V, 50 to 60 Hz, 70VA |
Dimensions (W × D × H) | 405 × 214 × 471 |
Packaging Dimensions (W × D × H) | 500 × 390 × 530 |
Net Weight | 13.13 kg |
Gross Weight | 16.73 kg |