Sample capacity | 96 wells |
Reaction volume | 10 to 50 μL |
Thermal cycle technology | Peltier |
Maximum Heating/Cooling rate | 6.0° C/s |
Heating temperature range | 4 to 100 °C |
Temperature accuracy | ± 0.2 °C |
Temperature uniformity | ± 0.2 ℃ at 60 ℃, ± 0.3 ℃ at 95 ℃ |
Temperature gradient setting range | 30 to 100 °C |
Temperature gradient difference setting range | 1 to 36 °C |
Excitation light source | 4 monochrome high efficiency LEDs |
Detector | PMT |
Detection mode | Time-resolved signal separating technology |
Excitation/detection wavelength range | 455 to 650 nm/510 to 715 nm |
Fluorescent channels | 5 channels |
Supported dye | FAM/SYBR Green, VIC/JOE/HEX/TET, ABY/NED/TAMRA/Cy3, JUN, ROX/Texas Red, Mustang Purple, Cy5/LIZ |
Sensitivity | Single copy gene |
Resolution | 1.33 folds copy number difference can be distinguished in single-plex qPCR |
Dynamic range | 10 orders of magnitude copies |
Packing Dimension (W × D × H) | 960 × 660 × 750 mm |
Gross weight | 47 Kg |