Multi Gas Analyzer FM-MGA-A101

Multi Gas Analyzer FM-MGA-A101

Multi Gas Analyzer FM-MGA-A101 rapidly examines concentration of 18 different gases. It features 10 step adjustable pump flow within range of 0 to 800 mL/min. 3 independent detection modules prevent cross interference and ensure accurate measurement. Incorporated with multiple sensors including electrochemical, IR and photoionization for enhanced sensitivity. Our portable and compact Gas Analyzer is essential for both qualitative and quantitative analysis.

  •  Multiple unit options
  •  Modular internal structure
  •  3.5-inch HD colour display
  •  Flexible display mode selection
  •  Wireless communication interface
  •  Screen capture functionality

Gases Detected O2, CH4, CO2, VOC, CO, H2S, SO2, Cl2, NO2, NH3, and more
Detection Range O2: 0 to 30 %Vol
CH4: 0 to 50000 ppm
CO2: 0 to 50000 ppm
VOC: 0 to 2000 ppm
CO: 0 to 999 ppm
H2S: 0 to 200 ppm
SO2: 0 to 25 ppm
Cl2: 0 to 50 ppm
NO2: 0 to 100 ppm
NH3: 0 to 100 ppm
Detection Resolution O2: 0.01 %Vol
CH4: 1 ppm
CO2: 1 ppm
VOC: 1 ppm
CO: 0.1 ppm
H2S: 0.1 ppm
SO2: 0.01 ppm
Cl2: 0.01 ppm
NO2: 0.01 ppm
NH3: 0.01 ppm
Concentration Units ppm, mg/m3, %Vol, %LEL
Working Principle Electrochemical (O2, CO, H2S, SO2, Cl2, NO2, NH3)
Infrared (CH4, CO)
Photoionization (VOC)
Detection Module 3 independent modules, each module detects 6 kinds of gases
Sampling Method Built-in flow pump, 1 in each of 3 modules
Pump Flow Type Adjustable in 10 steps
Pump Flow Range 0 to 800 mL/min
Temperature Range -20 °C to 50 ℃
Temperature Resolution 0.1 °C
Humidity Range 0 to 99% RH
Humidity Resolution 1% RH
Precision 2% FS
Response Time ≤ 10 s
Repeatability ≤ ± 1%
Zero Shift ≤ ± 1% (F.S / Year)
Recovery Time ≤ 10 s
Linearity Error ≤ ± 1%
Display Type 3.5-inch colour display
Display Resolution 320 × 480
Display Content Gas formula, Concentration data, Unit of measure, Gas curve trend, Battery level, Time, Temperature, Humidity, Pump status, Wireless status, Printer status, Screen capture status
Display Mode Single channel and Single curve mode, Multi channel data, Multi curve mode, Multi channel data and Multi curve mode
Working Environment Temperature -40 °C to 70 ℃
Working Environment Humidity 0 to 95% RH
Data Storage Standard 8G SD card storage with software
Communication Interface RS232, RS485
Wireless Transmission Type Bluetooth, ZigBee
Screen Capture Yes
Alarm Type Audio-visual, Alarm status prompt
Alarm Modes Fault alarm, Pump abnormal alarm, Low battery alarm, Shutdown alarm
Protection Level IP66
Explosion Proof Type Exic II CT3 Gc
Charging Interface Micro USB
Power Supply 3.7V DC, 10000 mA rechargeable polymer battery
Dimensions 260 × 230 × 120 mm
Weight 3.7 kg

Multi Gas Analyzer monitor hazardous gases, air quality index, pollutant and greenhouse gas levels in atmosphere.

Air Sampling Rod


Tube (600 mm)
Soot Filter
Charger
Carrying Case

Pressure Sensor


High Temperature Probe
Dry Filter
Printer