Light source | Xenon lamp 150 W |
Exciting optical system | Interference optical filter |
Optical filter bandwidth | 10 nm |
Optical filter beam size | 25 mm |
Central wavelength | 250 to 700 nm |
Emission monochromator | 200 to 900 nm (C-T diffraction grating) |
Detector | Multiplier tube |
Minimum sample volume | 1 mL |
Cuvette | Glass fluorescence cuvette 10 mm |
Wavelength accuracy | ±1 nm |
Wavelength repeatability | ≤ 0.5 nm |
Sensitivity | Raman peak of water in 1 cm quartz fluorescence cuvette with S/N ≥150 |
Linearity deviation | ≤ ±3.0 % |
Linear measurement | ≥ 0.995 |
Stability | 1.5% / 10 min |
Response time | (0.1 to 4) s with 6 stages adjustable |
Fluorescence display value | 0.00 to 600.00 |
Display | LCD digital display |
Standard interface | USB interface |
Working temperature | 15 to 30 ℃ |
Power supply | AC 110/220 V±10%, 60/50 Hz |
Dimension (L × W × H) | 550× 510× 365 mm |
Net weight | 12 kg |
Gross weight | 14 kg |