Double Beam UV-Visible Spectrophotometer FM-UVS-C200 is a compact, tabletop split double beam designed UV visible spectrophotometer comprising silicon photodiode detector, offers 190 to 1100 nm of wavelength range with 1.8 nm of spectral bandwidth. Features tungsten lamp and deuterium lamp as light source with dismountable lamp room cover, antirust treatment 4-cell manual holder with a special leak hole in sample room. Designed with 8 inch TFT color touch screen, 64 GB storage, this device have bluetooth wireless connected tablet PC with Windows operating system. Equipped with Optional software, USB-A and B, Wi-Fi, SD Reader, Bluetooth and HDMI interface and 10 languages, this device offers photometry, multi wavelength, kinetics, time scanning, quantitation, spectrum scanning, DNA/protein analysis along with data traceability.
Specifications
Wavelength range | 190 to 1100 nm |
Spectral bandwidth | 1.8 nm |
Optical system | Split beam; 1200 lines/mm grating |
Wavelength accuracy | ±0.3 nm |
Wavelength repeatability | ≤0.2 nm |
Wavelength resolution | 0.1 nm |
Wavelength swing speed | 10000 nm/min |
Wavelength scanning speed | 20 to 3200 nm/min |
Photometric range | 0 to 400% T, -4 to 4 A, 0 to 9999 C |
Photometric accuracy | ±0.3% T or ±0.004 A @ 1 A |
Photometric repeatability | ±0.001 A (0 to 0.5 A), ±0.002 A (0.5 to 1 A), ±0.15 % T (0 to 100 % T) |
Photometric Noise | ≤0.0002 A @ 0 A, 500 nm,≤0.0004 A @ 1 A,500 nm,≤0.0008 A @ 2 A, 500 nm |
Light source | Tungsten lamp, Deuterium lamp |
Stray light | ≤0.05 %T@ 220nm & 360nm |
Photometric Stability | ≤0.0005A/h (500nm, 1 hours after preheating) |
Display | 8 inch TFT color touch screen |
Baseline Flatness | ±0.001A |
Standard Sample holder | 10 mm 4-cell manual holder |
Detector | Silicon photodiode |
Function | Photometry, Multi Wavelength, Kinetics, Time Scanning, Quantitation, Spectrum Scanning, DNA/Protein analysis |
Storage | 64 GB |
Interface | USB-A (USB memory/USB printer), USB-B (computer), Wi-Fi, SD Reader, Bluetooth, HDMI |
PC software | Yes |
Languages | 10 Languages; Can Be Switched (English, German, French, Spanish, Portuguese, Italian, Russian, Simplified Chinese, Japanese, Korean) |
Power supply | AC 100 to 240 V,50/60 Hz,140 W |
Dimensions (W×D×H) | 456×360×185 mm |
Weight | 11 kg |
Features
Applications
Fison Double Beam UV-Visible Spectrophotometers provide high-precision optical system. They provide accurate and stable measurement of absorbance and transmittance. These units feature a double beam design, simultaneously measuring the sample and reference to reduce errors and improve accuracy. They include USB and printer connectivity, enabling seamless data transfer and efficient reporting. Our spectrophotometers offer a wide wavelength range for versatile analytical applications.
Fison Visible Spectrophotometer features a high-resolution LCD screen, offering clear and precise monitoring of parameters. They automatically set the wavelength, ensuring precise adjustments and simplifying the measurement process for accurate results. The large sample compartment fits cuvettes with various path lengths for efficient testing. Their data restoration after a sudden power cut ensures that important test results are not lost. Our Visible Spectrophotometer have grating optical system that ensures high spectral resolution.
Fison Single Beam UV-Visible Spectrophotometers are designed for accurate measurement of light absorbance and transmittance. They provide a wide wavelength range, covering the UV and visible spectrum for diverse applications. These units include automatic wavelength adjustment, ensuring precise and consistent wavelength selection. Our Spectrophotometers retain data after power loss, ensuring secure results. They provide USB and printer connectivity, allowing easy data transfer and efficient reporting.
Fison Fluorescence Spectrophotometers offer high sensitivity and selectivity, detecting low-concentration samples with superior accuracy. They are equipped with an advanced optical system, ensuring precise detection. These units provide automatic wavelength adjustment for accurate and consistent wavelength selection. They feature a user-friendly interface, offering a digital display and control system for easy operation. Our spectrophotometers come with a compact and durable construction for long-term reliability.
Fison Single Beam UV-Visible Spectrophotometers are designed for accurate measurement of light absorbance and transmittance. They provide a wide wavelength range, covering the UV and visible spectrum for diverse applications. These units include automatic wavelength adjustment, ensuring precise and consistent wavelength selection. Our Spectrophotometers retain data after power loss, ensuring secure results. They provide USB and printer connectivity, allowing easy data transfer and efficient reporting.
Fison Atomic Absorption Spectrophotometers offer high precision and sensitivity for detecting metal concentrations in various samples. They feature an advanced optical system with a high-resolution monochromator for enhanced spectral accuracy. These units provide automated sample handling, reducing manual intervention. They include a user-friendly interface and digital controls for easy operation. Our spectrophotometers deliver multiple calibration modes, allowing single and multi-element analysis to ensure versatility.
Wavelength range | 190 to 900 nm |
Wavelength accuracy | ±0.5 nm |
Resolution | Two spectral lines of Mn at 279.5 nm and 279.8 nm, can be separated with spectral bandwidth of 0.2 nm and valley-peak energy ratio less than 30% |
Baseline stability | 0.005 A/30 min |
Background correction | D2 lamp background correction capability at 1A is better than 30 times |
Lamp turret | 2 lamps powered simultaneously (one preheated) |
Lamp current adjustment | 0 to 20 mA |
Lamp power supply mode | 400 Hz square wave pulse |
Monochromator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 lines/mm |
Focal length | 277 mm |
Blazed wavelength | 250 nm |
Spectral bandwidth | 0.1 nm, 0.2 nm, 0.4 nm, 1.2 nm, automatic change |
Spectral adjustment | Manual adjustment for wavelength and slit |