Wavelength range | 190 to 1100 nm |
Spectral bandwidth | 1 nm |
Wavelength accuracy | ± 0.3 nm |
Wavelength repeatability | ≤ 0.2 nm |
Optical system | Double beam |
Photometric range | - 4 to 4 A, 0 to 400% T, 0 to 9999.9 T |
Photometric accuracy | ± 0.002 A at 0.0 to 0.5 A, ± 0.004 A at 0.5 to 1 A, ± 0.3 %T at 0 to 100 %T |
Photometric repeatability | ≤ 0.001 A at 0.0 to 0.5 A, ≤ 0.002 A at 0.5 to 1 A, ≤ 0.15 %T at 0 to 100 %T |
Detector | Dual Silicon Photodiodes |
Light source | Tungsten lamp, Deuterium lamp |
Stray light | ≤ 0.03 %T@ 220, 360 nm |
Work mode | A, % T, C |
Wavelength display | 0.1 nm |
Scanning speed | 20 to 3200 nm/min |
Noise | ≤ 0.0002 A at 0.0 A, 500 nm, ≤ 0.0004 A at 1 A, 500 nm,≤ 0.0008 A at 2 A, 500 nm |
Drift | ≤ 0.0005A/h at 500 nm, after warming up 1 hour |
Display | 10.1 Inch IPS color LCD with touch screen (1280 × 800) |
Baseline flatness | ≤ 0.0005 A |
Sample charger | 10 mm (Single) |
Storage | Built-in Storage 64 GB, Unlimited expansion (USB memory, SD card, network storage device) |
Communication | Interface USB-A, USB-B, RJ-45 (Ethernet), VGA, HDMI, Extensible Bluetooth, WIFI |
Power consumption | 140 W |
Power supply | 100 to 240 V AC,50/60 Hz |
Dimensions ( W × D × H ) | 580 × 420 × 235 mm |
Weight | 17 Kg |