Class II Biosafety Cabinet FM-BSC-A200 is designed with inward air flow system (negative air pressure) at a velocity (0.55 m/s) for personal protection, HEPA-filtered vertical laminar airflow for product protection, and HEPA-filtered exhausted air for environmental protection. Features motorized front window whose height can be adjusted using acquainted foot switch, remote control assessed operations and side waterproof sockets for operating small devices inside cabinet. Suitable for microbiological and research applications that do not involve radioactive substances and toxic contaminants.
Specifications
Type | Class II, Type B |
Filter Type | HEPA filter H14 |
Filter Quantity | 2 HEPA filters |
Filter efficiency | HEPA Filter: 99.995% efficiency at 0.3 µm |
Inflow velocity | 0.53 ± 0.025 m/s |
Down flow velocity | 0.33 ± 0.025 m/s |
UV lamp | 15 W ×1 |
UV Lamp emission | 253.7 nm |
Illuminating LED lamp | 8 W ×2 |
Illumination | ≥ 1000 Lux |
Power consumption | 500 W |
Noise level | ≤ 60 dB |
Waterproof socket | 500 W ×2 |
Front window | Two-layer laminated toughened glass, 5 mm |
Work surface height | 750 mm |
Tested Opening | Safety height ≤ 200 mm |
Max opening of front window | 370 mm |
Internal dimension (W×D×H) | 600×500×540 mm |
External dimension (W×D×H) | 700×650×1230 mm |
Display | LCD display |
Display parameter status | Status of cabinet, such as inflow and down flow velocity, work area temperature and humidity, filter pressure, UV working time and filter working time, filter life indicator. |
Control system | Microprocessor |
Airflow system | 70% air recirculation, 30% air exhaust |
Visual and audio alarm | Filter replacement, front window at unsafe height, abnormal air flow velocity |
Clean Level | ISO 5 Class 100 |
Chamber Material (work table) | 304 stainless steel |
Chamber Material (body) | Cold-rolled steel with anti-bacteria powder coating |
Power supply | 220 V±10%, 50/60 Hz, 100 V±10%, 60 Hz |
Packaging dimension with base stand (W×D×H) | 860×960×1450 mm |
Gross weight | 196 kg |
Features
Applications
Accessories
Accessories no | Accessories name |
1. | 2×UV lamp |
2. | LED lamp |
3. | Remote control |
4. | Waterproof Sockets |
Accessories Optional
Accessories No. | Name |
1 | Base Stand |
2 | Universal Caster with brake leveling feet |
Fison Class II Biosafety Cabinets offer personnel, product, and environmental protection by preventing contamination during biological sample handling. They are equipped with a laminar airflow design that maintains continuous airflow for optimal sample protection. Our cabinets feature a HEPA filtration system that effectively removes airborne contaminants. They provide negative pressure to prevent hazardous particle escape. These units deliver energy-efficient operation with optimized airflow and lighting for safety and performance.
Fison Class I Biosafety Cabinets offer personnel and environmental protection while handling low-to-moderate risk biological materials. They feature a HEPA filtration system that effectively removes airborne contaminants before air is released. These units are equipped with a negative airflow design that removes contaminated air, ensuring operational safety. They provide a functional workspace with a protective viewing panel for easy sample handling. Our cabinets include energy-efficient airflow and lighting to reduce power use while maintaining performance.
Fison Class III Biosafety Cabinets offer maximum containment protection for handling highly infectious and hazardous biological materials. They feature a gas-tight enclosure with a fully sealed workspace to prevent contamination. These units are equipped with dual HEPA filters for complete removal of hazardous particles. They include built-in heavy-duty gloves for secure material manipulation without direct exposure. Our cabinets deliver advanced safety alarms to monitor airflow, filter integrity, and pressure for maximum security.
Fison Cytotoxic Safety Cabinets offer a secure and sterile workspace for handling hazardous drugs and toxic compounds. They feature a HEPA filtration system that effectively removes airborne contaminants and drug residues. These units are equipped with a negative pressure design to prevent harmful substances from escaping into the environment. They provide laminar airflow technology to maintain a contamination-free workspace. Our cabinets deliver safety alarms and monitoring to detect airflow failures, filter status, and pressure changes.