Type | Class II, Type B |
Filter Type | HEPA filter H14 |
Filter Quantity | 2 HEPA filters |
Filter efficiency | HEPA Filter: 99.995% efficiency at 0.3 µm |
Inflow velocity | 0.53 ± 0.025 m/s |
Down flow velocity | 0.33 ± 0.025 m/s |
UV lamp | 15 W ×1 |
UV Lamp emission | 253.7 nm |
Illuminating LED lamp | 8 W ×2 |
Illumination | ≥ 1000 Lux |
Power consumption | 500 W |
Noise level | ≤ 60 dB |
Waterproof socket | 500 W ×2 |
Front window | Two-layer laminated toughened glass, 5 mm |
Work surface height | 750 mm |
Tested Opening | Safety height ≤ 200 mm |
Max opening of front window | 370 mm |
Internal dimension (W×D×H) | 600×500×540 mm |
External dimension (W×D×H) | 700×650×1230 mm |
Display | LCD display |
Display parameter status | Status of cabinet, such as inflow and down flow velocity, work area temperature and humidity, filter pressure, UV working time and filter working time, filter life indicator. |
Control system | Microprocessor |
Airflow system | 70% air recirculation, 30% air exhaust |
Visual and audio alarm | Filter replacement, front window at unsafe height, abnormal air flow velocity |
Clean Level | ISO 5 Class 100 |
Chamber Material (work table) | 304 stainless steel |
Chamber Material (body) | Cold-rolled steel with anti-bacteria powder coating |
Power supply | 220 V±10%, 50/60 Hz, 100 V±10%, 60 Hz |
Packaging dimension with base stand (W×D×H) | 860×960×1450 mm |
Gross weight | 196 kg |