Semi-Automatic Biochemistry Analyzer FM-SBC-A100 offers an absorbance range of 0.0–4.0 ensuring accurate analysis for both low and high-concentration samples. It uses back spectrophotometry to measure light absorption accurately for turbid or scattered samples. This unit features a separate signal processing circuit for each wavelength for multi-wavelength analysis. Our semi-automatic biochemistry analyzer adopts a touch screen for easy and accurate test result analysis.
Absorbance Range | 0.000 to 4.000 Abs |
Resolution | 0.001 Abs |
Temperature | 10℃ to 30℃ |
Humidity | 30 to 80% |
Temperature accuracy | ±0.1℃ at room temperature 37°C |
Temperature Control | Peltier elements |
Precision Error | <1% |
Analysis Method | Kinetics, End Point, Two Point, Multi-Point |
Optical System | 340 to 800 nm |
Standard Filters | 340, 405, 492, 510, 546, 578, 630 nm wavelength |
Filters Accuracy | ± 2 nm |
Flow Cell | Titanium quartz alloy |
Flow Cell Capacity | 33µL |
Optical Path | 10mm |
Precision Drift | <5mAbs/hr |
Printer | <1% |
CPU | Thermal printer, 57mm width paper |
Memory | Samsung S5PV210, Cortex-TM-AB, 1GHz |
Light Source | DDR2 SDRAM 256MB, Flash 4GB |
Interface | Halogen lamp 10W/6V |
Dimensions (L × W × H) | 3 USB ports, 1 RS232 serial port |
Net Weight | 380 × 280 × 170 mm |
Gross Weight | 5 Kg |
Applications:Semi-Automatic Biochemistry Analyzer FM-SBC-A100 performs precise biochemical tests including enzyme activity, glucose levels, and protein concentrations, and is used in clinical diagnostics, medical laboratories, and research institutions for routine tests and advanced biochemical analysis. |
Fison Instruments Ltd 272 Bath Street Glasgow G2 4JR UK
Email: info@fison.com | Website: www.fison.com