Description :

Melting Point Apparatus FM-MP-A100 is a desktop apparatus used to determine the melting point of a substance. Improved with automated technology measures powdered materials with a melting point temperature range of 280°C. Featured a user-friendly display with a precision of 0.1°C. Equipped with silicon Heating conducting medium. Functionality is crucial in identifying and characterizing various organic and inorganic compounds that have distinct melting points.


Features:

  • Engineered to hold the sample securely and promote consistent melting
  • Supplies a sample tube up to an external diameter of 1.4 mm and an internal diameter of 1.0mm
  • Provides a 3 mm loading height for samples and can measure three samples
  • Automates the calculation of average value for both the initial and final melting points
  • Utilizes an automatic detection method coupled with a cooling function

Specifications:


Measuring Range 280◦C
Accuracy ≤ 200˚C, ± 0.4˚C, 200˚C, ±0.7˚C
Repeatability 0.3◦C
Heat Conductive medium Silicon oil
Size of the capillary Length:120mm
Outside Diameter: Φ 1.4mm
Inside Diameter: Φ 1.0mm
No. of Capillary 3
Loading height of sample 3mm
Minimum Digital Display 0.1◦C
Interface RS232
Dimensions 390 × 390 × 520mm
Weight 13kgs

Applications:

It is applied in various industries, including drug industry, biochemistry, medicines, mining, and water management, for determining the melting points of medicines, dyestuffs, perfumes, and other organic crystals.


www.fison.com

Fison Instruments Ltd 272 Bath Street Glasgow G2 4JR UK

Email: info@fison.com | Website: www.fison.com