Handheld Spectrophotometer FM-HSP-A100 has dual optical path system. It uses LED light source with long life and low power consumption. The spectral range is of about 400 to 700 nm and wavelength pitch of 10 nm. Equipped with sturdy and stable white calibration board. The great storage capacity of the product makes it stand out in the competition. Meter’s expandable functionality makes it a versatile tool in the testing of solutions.
Optical Geometry | Reflect: di:8°, de:8° (diffused illumination, 8-degree viewing angle) |
Integrating Sphere Size | Φ40 mm |
Light Source | Combined full spectrum LED light source, UV light source |
Spectrophotometric Mode | Flat Grating |
Sensor | Silicon photodiode array (double row 40 groups) |
Wavelength Range | 400 to 700 nm |
Wavelength Interval | 10 nm |
Semi band Width | 10 nm |
Measured Reflectance Range | 0 to 200% |
Measuring Aperture | MAV: Φ8mm/Φ10mm;SAV: Φ4mm/Φ5mm |
Specular Component | SCI&SCE |
Color Space | CIE LAB, XYZ, Yxy, LCh, CIE LUV, s-RGB, HunterLab, βxy, DIN Lab99 Munsell(C/2) |
Color Difference Formula | ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc (2:1),ΔE*cmc (1:1),ΔE*00, DINΔE99,ΔE(Hunter) |
Other Colorimetric Index | WI(ASTM E313,CIE/ISO,AATCC, Hunter, YI (ASTM D1925,ASTM 313), Metamerism Index MI, Staining Fastness, Color Fastness, Color Strength, Opacity, 8° Glossiness,555 tone classification |
Observer Angle | 2°/10° |
Illuminant | D65, A, C, D50, D55, D75, F1, F2(CWF), F3, F4, F5, F6, F7(DLF), F8, F9, F10(TPL5), F11(TL84), F12(TL83/U30) |
Data display | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Offset |
Measuring Time | About 1.5s (Measure SCI & SCE about 3.2s) |
Repeatability |
Spectral reflectance: MAV/SCI, Standard deviation within 0.08% (400 nm to 700 nm: within 0.18%) Chromaticity value: MAV/SCI, within ΔE*ab 0.02 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Inter-instrument Error | MAV/SCI, Within ΔE*ab 0.15(Average for 12 BCRA Series II color tiles) |
Measurement Mode | Single Measurement, Average Measurement (2 to 99 times) |
Locating Method | Camera Locating, stabilizer cross position |
Battery | Li-ion battery, 6000 measurements within 8 hours |
Illuminant Life Span | 5 years, more than 3 million times measurements |
Display | 3.5-inch TFT color LCD |
Communication | USB, Bluetooth |
Data Storage | Standard 1000 Pcs |
Operating Environment | 0 to 40℃, 0 to 85%RH (no condensing), Altitude < 2000 m |
Storage Environment | -20 to 50 ℃, 0 to 85%RH (no condensation) |
Dimension (L × W × H) | 129 × 76 × 217 mm |
Packing dimensions (L × W × H) | 370 × 310 × 170 mm |
Net Weight | 0.6 Kg |
Gross Weight | 3 Kg |
Applications:Handheld Spectrophotometer is solely used across textiles, environmental monitoring, foods and beverages, paints and coatings, agriculture, etc. |
Accessories Optional:Micro Printer |
Fison Instruments Ltd 272 Bath Street Glasgow G2 4JR UK