Our specialists will reach out shortly.

Flame Atomic Absorption Spectrometer FM-FAAS-A100 Catalog

Overview

Flame Atomic Absorption Spectrometer FM-FAAS-A100 with a wavelength range of 190 to 900 nm detects various elements accurately. It has an adjustable spectral bandwidth of 0.1, 0.2, 0.4, 1.0, and 2.0 nm for optimal resolution and sensitivity. It features dual background correction function to reduce spectral interferences and improve measurement accuracy. Our spectrometer has an optical visual system that ensures clear sample observation for accurate analysis.

Specifications :

Wavelength Range 190 to 900 nm
Spectral Bandwidth 0.1, 0.2, 0.4, 1.0, 2.0 nm (automatically adjustable)
Temperature +15°C to +35°C
Relative Humidity 20 to 80%
Wavelength Accuracy <0.15 nm
Wavelength Repeatability ±0.1 nm
Baseline Stability <±0.002A/30 minutes (static), <±0.005A/30 minutes (dynamic)
Light Source 8-lamp automatic turret with automatic alignment
Power Supply for Lamps Double cathode power built-in high-performance lamps
Characteristic Concentration (Cu) 0.015 pg/mL/1%
Detection Limit (Cu) 0.002 pg/mL
Precision RSD = 0.5%
Combustion Head Metal titanium combustion head
Atomizer Efficient glass atomizer
Atomizing Chamber Explosion-proof, corrosion-resistant spray chamber
Control System Automatic PC control, three-light brick alignment, automatic optimization, automatic ignition
Safety Features Automatic safety protection, anti-tempering gas path protection, acetylene gas leakage alarm, automatic shutdown system, abnormal power-off protection
Background Correction Deuterium background correction (correction for 1A background)
Measurement Methods Flame method, hydride method
Concentration Calculation Methods Standard curve method (1-3 times curve), automatic matching, standard addition method
Repetition Survey Frequency 1 to 99 times, with calculation of average value, standard deviation, and relative standard deviation
Results Printing & Export Parameter print, data and graphics print, export to Word/Excel
Operation Simple software-controlled operation, automatic lamp position rotation, automatic ignition
Communication Interface Computer and USB interface
Power Supply 5000VA
Voltage 220V (+5% to -10%), 50/60 Hz
Dimension (L × W × H) 99 cm × 43 cm × 41 cm
Packaging Dimension (L × W × H) 126 × 64 × 71 cm
Net Weight 100 Kg
Gross Weight 120 Kg

Features :

  • Double Cast Aluminum Plate
  • Burner Deflection Control Technology
  • High-Precision Lead Screw Mechanism
  • High-Performance Dual Cathode Hollow Lamp
  • Precision Flow Controller
  • High-Precision Rotameter
  • High-Sensitivity Photomultiplier

Applications :

  • Flame Atomic Absorption Spectrometer is ideal for Environmental Analysis, Food and Beverage Industries, Pharmaceutical Industries, Clinical and Biomedical Applications, Petroleum and Petrochemical Industries.
  • Accessories :

    1.Desktop computer (1 set)

    2. Printer (1 set)
    3. Oil-less Air Compressor (1 set)
    4. Hollow cathode lamp (3 pcs)
    5. Atomizer (2 pcs)
    6. Spare part Carton (1 box)
    7. Main unit dust cover (1 pc)
    8. 3A fuse wire (1 pc)

    9. 1.5A fuse wire (1 pc)

    10. Three core power line (1 pc)

    11. Main Unit Signal line (1 pc)
    12. Plastic pipe ф6mm (5meters), ф10mm (1.5meters ) (2 pcs)

    13. Screwdriver (1 pc)

    14. Acetylene hose clip (3 pcs)

    15. Acetylene gas pipe (3 meter)

    16.  HCL Lamp (Cu, Pb, Cd) (1 pc)

    Accessories Optional :

    1.  HCL Lamp (Au, Ag, Ni, Co, Fe, Cr, Mn, Li, Hg, Zn)

    2.  Auto Sampler (Flame)

    3.  Hydride generator 

    4.  Cooling water System 

    5.  Glass Atomizer 

    6.  Adjustable atomizer (1-10mL) 

    7.  Hydrofluoric acid resistant type atomizer 

    8.  Electrically heated quartz tube (normal) 

    9.   Electrically heated quartz tube (White porcelain)

    Fison