Automatic Potentiometric Titrator FM-APT-A101 is incorporated with 4 channels for high-throughput processing. Microprocessor controlled, fully automated operation facilitates reproducible and reliable titrations. Equipped with pH and mV electrodes for dual mode measurement. Features audit trail and advanced password settings for streamlined data documentation and security. It determines sample endpoints as per ASTM D664 and D2896 industry protocols.
Titration Types |
Potentiometric titration Polarization current titration |
Measurement Modes | Measurement Mode (MEAS: pH, mV, T), Calibration Mode (CAL: pH), Endpoint Setting Mode (SET), Method Mode (MET: Equal Volume Titration, Second-Order Derivative Chloride Ion Measurement), Dynamic Titration Mode |
Analysis Methods | Acid-base, Prescription, Non-aqueous, Complexometric, Permanent-stop titrations |
Measurement Range |
0.01 mg to 50 mg of KOH / g mV: (-2000 to 2000) mV pH: (-20 to 20) pH Temp: (-5 to 120) °C |
Measurement Accuracy |
± 0.02 mg of KOH / g mV: 0.1 mV ± 0.03 % pH: 0.003 pH Temp: ± 0.1°C |
Resolution |
mV: 0.1 mV pH: 0.001 pH |
Burette Capacity | 10 mL standard burette (1/5/20 mL are optional) |
Burette Refilling Time | 16 seconds |
Burette Resolution | 1/1500000 |
Number of Titration Channels | 4 |
Standard Temperature | 20°C ± 5°C |
Operating Environment Temperature | 10°C to 35°C |
Relative Humidity | 50% -70% |
Interface | USB, RS232, Ethernet |
Power Supply | AC 220V ± 10%, 50 Hz |
Packaging Dimensions | 30 × 25 × 20.5 cm |
Net Weight | 14 kg |
Gross Weight | 15 kg |
Applications:Automatic Potentiometric Titrators aid in quality control and testing of pharmaceuticals, chemicals, cosmetics, petroleum products, food and beverages. |
Accessories:
|
Fison Instruments Ltd 272 Bath Street Glasgow G2 4JR UK
Email: info@fison.com | Website: www.fison.com