Atomic Absorption Spectrophotometer FM-AAS-A201 is a flame type automatic PC controlled analyzer. Unit is assembled with 4-lamp turret, air-acetylene flame, D2 background correction, software and Cu, Mn, Hg hollow cathode lamps. Adopted with rich oxygen air-acetylene flame analysis technique as substitution to nitrous oxide-acetylene flame for high temperature element analysis such as Ca, Al, Ba, W, Mo, Ti, V, etc. Flame temperature is continuously adjustable between 2300 to 2950°C, enables it to select best atomization temperature for different elements. Features easy operation, low cost analysis and wide flame AAS analytical range. Its Rich oxygen flame is environmental nonpolluting and harmless to human bodies.
Wavelength Range | 190 to 900 nm |
Wavelength Accuracy | ± 0.25 nm |
Resolution | Two spectral lines of Mn at 279.5 nm and 279.8 nm, separable with a spectral bandwidth of 0.2 nm and valley-peak energy ratio < 30% |
Baseline Stability | ≤0.004 A/30 min |
Background Correction | D₂ lamp background correction, capability at 1A is better than 30 times |
Lamp Turret | 4-lamp turret auto-alignment, fully automated scan and peak-picking |
Lamp Current Adjustment | Automatic adjustment and display |
Wide pulse current | 0 to 25 mA |
Narrow pulse current | 0 to 10 mA |
Lamp Power Supply Mode | 400 Hz square wave pulse |
Monochromator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 lines/mm |
Focal Length | 277 mm |
Blazed Wavelength | 250 nm |
Spectral Bandwidth | 0.1 nm, 0.2 nm, 0.4 nm, 1.2 nm (automatic change) |
Burner | 10 cm single-slot all-titanium burner |
Spray Chamber | Corrosion-resistant all-plastic spray chamber |
Nebulizer | High-efficiency glass nebulizer with metal sleeve, sucking up rate: 6 to 7 mL/min |
Detector | R928 Photomultiplier with high sensitivity and wide spectral range |
Result Printing | Measurement data and final analytical report printout, editable in Excel |
Software | Windows operating system |
Analytical Method | Working curve auto-fitting, standard addition method, automatic sensitivity correction, automatic concentration and content calculation |
Repeat Times | Maximum 20 repeat measurements, automatic calculation of mean, standard deviation, and relative standard deviation |
Multi-task Function | Sequential measurement of multiple elements in one sample |
Condition Reading | Model function available |
Standard Interface | Standard RS232 serial port communication |
Characteristic Concentration of Cu | ≤0.025 mg/L |
Detection Limit of Cu | ≤0.006 mg/L |
Power Supply | AC 220 V, 50 Hz |
Dimension | 122 × 66 × 89 mm |
Weight | 160 kg |
Used in clinical analysis of metals in biological fluids and tissues such as whole blood, plasma, urine, saliva, brain tissue, liver, hair, muscle tissue, semen, in some pharmaceutical manufacturing processes, minute quantities of a catalyst that remain in final drug product, and analyzing water for its metal content.