Carbon Measuring Range | 0.0005% to 6.0000% (can be extended to 99.999%) |
Sulfur Measuring Range | 0.0005% to 0.5000% (can be extended to 99.999%) |
Sample weight | Standard 0.5g |
Carbon Analysis Error | In accordance with ISO9556 standard |
Sulfur Analysis Error | In accordance with ISO4935 standard |
Carbon Analysis Precision | RSD<1% |
Sulfur Analysis Precision | RSD<1.5% |
Carbon Analysis pool | One pool |
Sulfur Analysis pool | One pool |
Time of analysis | 25 to 60 seconds could be adjusted. Around 35 seconds usually |
Sensitivity | Minimum reading of 0.1 ppm for carbon/sulfur (C/S). |
Precision of read | 0.0001g |
Room Temperature | 10 to 30 degree C |
Relative Humidity | <75% |
Power Supply | 220v 50hz |
Dimensions | 138 × 76 × 98 mm |
Weight | 192 Kg |