Sample | Whole blood, Serum, Plasma, Dialysate, CSF |
Sample Volume | 120 µl whole blood , 50 µl capillary blood |
Analysis Time | < 90s |
Test Parameter | pO2, pCO2, pH, K, HCT.B.P, Na+ , Cl- , Ca++ |
Operating Temperature | 15 to 30°C |
Storage Environment | 5 to 45°C |
Input Variable | Patient temperature, Hemoglobin, FiO2 , Barometric Pressure |
Measured Factor |
Hydrogen ion activity (pH): 6.000 to 9.000 pH Partial pressure (pCO2): 1.607 to 26.67 Kpa, 8.0 to 200.0 mmHg (PO2): 0 to 106.7 kPa,0 to 800.0 mmHg Potassium ion (K): 0.5 to 15.0 mmol/L Sodium ion (Na+): 20.0 to 200.0 mmol/L Chloride ion (Cl-): 20.0 to 200.0 mmol/L Ionized calcium (Ca++) : 0.3 to 5.0 mmol/L Hct: 12.0 to 65.0% B.P. : 500 to 800 mmol/L |
Derived Factors | pH (CH)pCO2 , pO2 , K , Na+ , Cl- , Ca++ , Hct , B.P |
Calibration | Automatic, programmable 1-point and 2-point calibration |
Data Storage | > 5000 result |
Display | 10.4 inches TFT touch screen |
Printer | Built-in thermal printer |
Bar Code Reader | Multiple types barcode type |
Interface | Serial line RS-232, TCP/IP interface, USB port, Integrated Bar-Code Reader |
Battery | 24 Nickel-metal hydride battery, 3.0 Ah |
Power Supply | 100 to 240 V, 50/60 Hz |
Reagent Dimension (W × D × H ) | 586 × 460 × 380 mm |
Packaging Dimension (W × D × H ) | 920 × 650 × 610 mm |
Gross Weight ( Reagent ) | 9.7 kg |
Gross Weight (Main Instrument ) | 52.8 kg |