- ✔ Integrated flame/graphite furnace atomization system, changeable with flame emission burner
- ✔ Turret for 6 pre-aligned hollow cathode lamps fitted with pre-heating, and individual pre-centered lamps
- ✔ Spectral bandwidth of 0.2 nm, 0.4 nm, 0.7 nm, 1.4 nm, 2.4 nm, 5.0 nm
- ✔ Characteristic concentration of copper: ≤ 0.04 mg/ml/1%
- ✔ Detection limit of copper: ≤0.008 mg/ml
- ✔ Built-in computer data processing system and LCD display
- ✔ Double-beam system compensates light source and wavelength drift caused by temperature variation
- ✔ Gas path system equipped with precision pressure stabilizing and current stabilizing devices
- ✔ Multi-functional analysis mode: flame absorption, flame emission, graphite furnace atomic absorption and hydride generation
- ✔ Integrated functions: integral holding, peak height and area, auto zero adjusting, deuterium lamp background correction, multi-linear and nonlinear curves fitting
- ✔ Various parameters and working curves displayed in screen and report printing
- ✔ It is equipped with autosampler and RS232 interface for externally linking PC
Wavelength range |
190 to 900 nm |
Wavelength accuracy |
≤ ±0.5 nm |
Wavelength repeatability |
≤0.3 nm (single direction) |
Spectrum bandwidth |
0.2 nm, 0.4 nm, 0.7 nm, 1.4 nm, 2.4 nm, 5.0 nm |
Resolution |
< 40% |
Base line stability |
±0.004 Abs/30 min |
Grating |
1200 lines / mm |
Inflamer |
All-metal titanium burner with 10 cm slit |
Atomizer |
Efficient glass atomizer |
Lamp |
Hollow cathode lamp |
Lamp stand |
6 lamps automatic turret, automatic alignment |
Characteristic concentration of copper |
≤0.04 mg/ml/1% |
Detection limit of copper |
≤0.008 mg/ml |
D2 Background calibration ability |
Greater than 30 times |
Flame system |
Air-acetylene flame system |
Protection system |
Can inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner |
Standard interface |
RS232 |
Power supply |
220 V/ 50 Hz, 3 A |
Gross weight |
160 kg |
Used in different areas of chemistry such as clinical analysis of metals in biological fluids and tissues such as whole blood, plasma, urine, saliva, brain tissue, liver, hair, muscle tissue, semen, in some pharmaceutical manufacturing processes, minute quantities of a catalyst that remain in the final drug product, and analyzing water for its metal content.
Accessories no |
Accessories name |
1. |
Oil-free air compressor |
2. |
Glass Atomizer |
3. |
Cu Hollow cathode lamp |
4. |
Atomizer unit |
5. |
Burner unit |
6. |
Dust cover |
7. |
Water-separating gas filter |
8. |
Titanium burner: slit 10 cm |
Accessories no |
Accessories name |
1 |
GRAPHITE FURNACE SYSTEM:
|
|
Heating steps- 9 steps |
|
Temperature range: 20°C to 3000°C |
|
Slope heating time: 0 to 999 s |
|
Heating holding time: 0 to 999 s |
|
Inert gas: argon |
|
Inert gas pressure: 0.3 MPa |
|
Cooling water flow rate: 2 L/min |
|
Characteristic quantity: (Cd) ≤ 1 × 10 -12 g, (Cu) ≤ 1 × 10 -10 g |
|
Power supply: 220 V/50 Hz 3A |
|
Ambient temperature: 10 to 30 °C |
|
Relative humidity: less than 85% |
2 |
Hydride generator |
3 |
Graphite tubes |
4 |
Acetylene Regulator |
5 |
Printer |
6 |
Software |